PHG-11TZ+/PHG-12TZ+

Surge Protection Type RTD input temperature converter

PHG-11TZ+  1 input 1 output
PHG-12TZ+  1 input 2 outputs

Input :Two-wire or three-wire RTD signal

Output :DC signal (current/voltage)

Request a Quote

PRODUCTS DETAILS

Overview

The RTD input temperature converter with surge protection function can convert the two wire or three wire thermistor (RTD) signals into DC current signals. It can be intelligently configured, and the actual range of the thermal resistance can be set through computer software. It has the functions of wire breakage alarm and out of range alarm.

This product requires independent power supply, with isolated power supply, input, and output terminals.

 

Technical Data

Input
Input signal Two wire or three wire thermistor signals (see "Input Signal Type and Range Table" for details)
Input disconnection The default "low alarm" can be modified to "high alarm" through configuration software
Signal range Corresponding measurement range of thermal resistance
Measurement range Users make their own configuration when ordering, and indicate it at the tail number or otherwise.
Excitation current 800μΑ
Output:
Output signal DC signal (current/voltage)
Output load capacity
Surge protection features:
Nominal discharge current ln(8/20μs) 5kA
Voltage protection level Up(8/20μs): 60V(line to line)
Voltage protection level Up(8/20μs): 600V(line to ground)
According to standards GB/T18802.21-2016(equivalent to 1EC61643-21:2012)
Power consumption Current output <0.8W; voltage output <0.6W (PHG-11TZ+)
Current output <1.2W; voltage output <0.6W (PHG-12TZ+)
LED indicator  Green: Power indicator
Low range alarm L1 light on, high range alarm L2 light on
Output accuracy Please refer to the "Input Signal Type and Range Table" for details
Response time Reaching 90% of the final value within 300ms
Temperature drift 0.005%F.S/℃
Temperature parameters Working temperature: -20℃~+60℃, storage temperature:-40℃~+80℃
Relative humidity 10%~95% RH no condensation
Dielectric strength ≥ 2000VAC/min (between input/output/power supply)
Insulation resistance 100MΩ (500VDC)(between input/output/power supply)
Electromagnetic compatibility GB/T 18268(IEC 61326-1)
MTBF 80000h
Wire requirements Horizontal cutting surface ≥ 0.5mm2; Insulation strength ≥ 500V
Applicable field equipments Two wire or three wire thermistors
G53, Cu50, Pt100, Pt1000, Ni1000

Dimension

PHG-11TZ+.jpg

Schematic diagram

PHG-11TZ+(1).jpg

Note:

1. PHG-11TZ+ does not include output part 2

2. The power rail function is an optional function, and users need to specify the power supply method when placing an order

The selection of power rail connectors can refer to page 89 of the "Annex"

3. When inputting a three wire RTD, it is necessary to ensure that the three wires are of equal length as much as possible

4. When inputting a two-wire RTD, safety barrier terminals 4 and 2 must be short circuited

 

Input signal types and measurement range
Code RTD model Measurement range  Minimum range Conversion accuracy
1 G53 -50~150℃ 20℃ 0.2℃/0.1%
2 Cu50 -50~150℃ 20℃ 0.2℃/0.1%
4 Pt100 -200~850℃ 20℃ 0.2℃/0.1%
6 Pt1000 -200~850℃ 20℃ 0.2℃/0.1%
7 Ni1000 -60~250℃ 20℃ 0.2℃/0.1%
Output parameter definition
Code Parameter
1 4~20mA
2 0~20mA
3 0~5V
4 0~10V
5 1~5V
6 0.042361111
7 土10V
8 customized

PHG-11TZ+(2).jpg

For example: one input and two outputs, RTD Pt100 (-200~850°C) input, two channels 4~20mA outputs, 24V power supply, model: PHG-12TZ-411+

Contact us

Please feel free to give your inquiry in the form below We will reply you in 24 hours

Leave YourMessage

Thank you! We will reply you via email within 24 hours.

return